Last checked: Nov 19, 2026Latest source record: Oct 30, 2026 (newest report received)openFDA last updated: Nov 11, 2026Methods & limitations
Filters
Filters are saved in the page address, so a filtered view can be bookmarked or shared.
With no dates set, the view covers the last 12 full months and the current month. Brand and manufacturer are matched as FDA recorded them, so try the short form of a name. Product code QBJ: Integrated Continuous Glucose Monitoring System, Factory Calibrated. Open this device type's page
Reports received
636,709
Received Nov 1, 2025 to Nov 19, 2026. Includes 559 of another or unstated type.
Malfunction reports
617,141
97% of reports
Injury reports
18,855
3.0% of reports
Death reports
154
0.0% of reports
Reports received per month
product code QBJ. By the date FDA received each report, stacked by event type. The newest month is incomplete, and FDA adds reports for recent months as it processes them.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Nov 25 | 23,245 | 1,015 | 14 | 0 |
| Dec 25 | 26,379 | 1,475 | 20 | 0 |
| Jan 26 | 82,421 | 1,280 | 25 | 5 |
| Feb 26 | 21,874 | 1,398 | 13 | 0 |
| Mar 26 | 27,589 | 1,330 | 14 | 2 |
| Apr 26 | 52,802 | 1,781 | 4 | 0 |
| May 26 | 57,186 | 1,845 | 9 | 0 |
| Jun 26 | 65,787 | 1,789 | 14 | 0 |
| Jul 26 | 48,681 | 1,930 | 11 | 55 |
| Aug 26 | 92,354 | 1,651 | 13 | 122 |
| Sep 26 | 81,854 | 1,888 | 6 | 140 |
| Oct 26 | 36,969 | 1,473 | 11 | 235 |
| Nov 26 | 0 | 0 | 0 | 0 |
Device types with the most reports
By FDA product code. A report that names several devices counts once for each.
Manufacturers with the most reports
Bigger sellers get more reports. This is not a ranking of safety.
Brand names with the most reports
As written on the reports. One product can appear under several spellings.
- FREESTYLE LIBRE 3124,533
- FreeStyle Libre 3 Plus112,412
- FreeStyle Libre 2 Plus65,380
- DEXCOM G7466
636,709 reports
Showing 1 to 25, newest first. Each title opens FDA's full record, which includes the reporter's description of the event.
- FreeStyle Libre 3 PlusMalfunction
2954323-2026-448743 · Received Oct 30, 2026 · Event dated Oct 19, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Device Alarm System
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- FreeStyle Libre 2 PlusInjury
2954323-2026-448742 · Received Oct 30, 2026 · Event dated Oct 5, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: High Readings
Patient problem as coded: Hyperglycemia
3004753838-2026-289462 · Received Oct 30, 2026 · Event dated Oct 14, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289464 · Received Oct 30, 2026 · Event dated Oct 7, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289468 · Received Oct 30, 2026 · Event dated Oct 14, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289479 · Received Oct 30, 2026 · Event dated Sep 29, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289483 · Received Oct 30, 2026 · Event dated Oct 10, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289493 · Received Oct 30, 2026 · Event dated Oct 11, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289480 · Received Oct 30, 2026 · Event dated Oct 12, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289490 · Received Oct 30, 2026 · Event dated Oct 13, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289491 · Received Oct 30, 2026 · Event dated Oct 2, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289459 · Received Oct 30, 2026 · Event dated Oct 20, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289470 · Received Oct 30, 2026 · Event dated Oct 13, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Low Readings
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289482 · Received Oct 30, 2026 · Event dated Oct 15, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Unintended Application Program Shut Down
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289473 · Received Oct 30, 2026 · Event dated Jun 15, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: No Device Output
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289460 · Received Oct 30, 2026 · Event dated Oct 18, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Appropriate Device Problem Term/Code Not Available
Patient problem as coded: Hypoglycemia; Ambulation or Postural Difficulties
3004753838-2026-289465 · Received Oct 30, 2026 · Event dated Jun 25, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: No Device Output
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289467 · Received Oct 30, 2026 · Event dated Apr 1, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289474 · Received Oct 30, 2026 · Event dated Oct 14, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289488 · Received Oct 30, 2026 · Event dated Oct 15, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Unintended Application Program Shut Down
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289489 · Received Oct 30, 2026 · Event dated Oct 21, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289461 · Received Oct 30, 2026 · Event dated Oct 18, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289484 · Received Oct 30, 2026 · Event dated Oct 16, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Wireless Communication Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289457 · Received Oct 30, 2026 · Event dated Oct 10, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Incorrect, Inadequate or Imprecise Result or Readings
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
3004753838-2026-289466 · Received Oct 30, 2026 · Event dated Oct 17, 2026
Device type: Integrated Continuous Glucose Monitoring System, Factory Calibrated (QBJ)
Device problem as coded: Low Readings
Patient problem as coded: Hyperglycemia
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.