Device type · NKE
Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
NKE
Cardiovascular
Device class
Class III
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
3,073
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)
- Product code
- NKE
- Device class
- Class III
- Review panel
- Cardiovascular
- Specialty
- Cardiology, Electrophysiology
- FDA record
- Classification database entry for NKE
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
3,073
Received Nov 1, 2025 to Nov 19, 2026. Includes 4 of another or unstated type.
Malfunction reports
1,717
56% of reports
Injury reports
1,338
44% of reports
Death reports
14
0.5% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 53 | 74 | 0 | 0 |
| Jan 24 | 57 | 90 | 2 | 0 |
| Feb 24 | 50 | 62 | 3 | 0 |
| Mar 24 | 44 | 73 | 2 | 0 |
| Apr 24 | 64 | 68 | 3 | 0 |
| May 24 | 59 | 69 | 3 | 0 |
| Jun 24 | 57 | 56 | 0 | 0 |
| Jul 24 | 58 | 63 | 1 | 0 |
| Aug 24 | 53 | 72 | 2 | 0 |
| Sep 24 | 54 | 70 | 3 | 0 |
| Oct 24 | 58 | 77 | 1 | 0 |
| Nov 24 | 46 | 75 | 0 | 0 |
| Dec 24 | 60 | 61 | 2 | 0 |
| Jan 25 | 44 | 85 | 0 | 0 |
| Feb 25 | 54 | 85 | 3 | 0 |
| Mar 25 | 67 | 87 | 2 | 0 |
| Apr 25 | 64 | 75 | 5 | 0 |
| May 25 | 62 | 78 | 2 | 0 |
| Jun 25 | 44 | 68 | 2 | 0 |
| Jul 25 | 52 | 64 | 1 | 0 |
| Aug 25 | 43 | 63 | 2 | 0 |
| Sep 25 | 60 | 90 | 2 | 0 |
| Oct 25 | 82 | 168 | 4 | 0 |
| Nov 25 | 96 | 130 | 0 | 0 |
| Dec 25 | 93 | 178 | 2 | 0 |
| Jan 26 | 94 | 165 | 0 | 0 |
| Feb 26 | 89 | 158 | 3 | 0 |
| Mar 26 | 154 | 131 | 1 | 0 |
| Apr 26 | 168 | 138 | 2 | 0 |
| May 26 | 149 | 107 | 2 | 0 |
| Jun 26 | 178 | 74 | 0 | 0 |
| Jul 26 | 222 | 79 | 1 | 0 |
| Aug 26 | 117 | 57 | 0 | 1 |
| Sep 26 | 181 | 59 | 2 | 1 |
| Oct 26 | 176 | 62 | 1 | 2 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Boston Scientific2,339
- Medtronic481
- Abbott208
- Unk2
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Premature Discharge of Battery1,641
- Adverse Event Without Identified Device or Use Problem338
- Interrogation Problem307
- Incorrect, Inadequate or Imprecise Result or Readings227
- Over-Sensing187
- Pacing Problem183
- Use of Device Problem157
- Data Problem116
- Signal Artifact/Noise99
- Failure to Capture97
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions2,493
- Unspecified Infection225
- Syncope/Fainting46
- Insufficient Information39
- Medical device site infection39
- Discomfort34
- Bradycardia33
- Pocket Erosion30
- Asystole28
- Device Overstimulation of Tissue28
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1771-2026 · Started Mar 19, 2026 · Posted by FDA Apr 24, 2026
Reason given: Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are...
Cause as FDA categorized it: Nonconforming Material/Component
- Model Number U125, VALITUDE CRT-P ELOpen, Classified
Z-0101-2026 · Started Aug 20, 2025 · Posted by FDA Oct 10, 2025
Reason given: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life...
Cause as FDA categorized it: Other
- Model Number U128, VALITUDE CRT-P EL MRIOpen, Classified
Z-0102-2026 · Started Aug 20, 2025 · Posted by FDA Oct 10, 2025
Reason given: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life...
Cause as FDA categorized it: Other
- Model Number U225 VISIONIST CRT-P ELOpen, Classified
Z-0103-2026 · Started Aug 20, 2025 · Posted by FDA Oct 10, 2025
Reason given: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life...
Cause as FDA categorized it: Other
- Model Number U226, VISIONIST CRT-P ELOpen, Classified
Z-0104-2026 · Started Aug 20, 2025 · Posted by FDA Oct 10, 2025
Reason given: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life...
Cause as FDA categorized it: Other
- Model Number U228, VISIONIST CRT-P EL MRIOpen, Classified
Z-0105-2026 · Started Aug 20, 2025 · Posted by FDA Oct 10, 2025
Reason given: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life...
Cause as FDA categorized it: Other
Z-0826-2025 · Started Dec 12, 2024 · Posted by FDA Jan 10, 2025
Reason given: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
Cause as FDA categorized it: Process design
Z-0827-2025 · Started Dec 12, 2024 · Posted by FDA Jan 10, 2025
Reason given: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
Cause as FDA categorized it: Process design
Showing the 8 newest of 28 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.