Device type · DXE
Catheter, Embolectomy
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
DXE
Cardiovascular
Device class
Class II
FDA decisions since 2015
11
510(k), De Novo and PMA
Adverse event reports
499
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Catheter, Embolectomy
- Product code
- DXE
- Device class
- Class II
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5150
- Specialty
- Cardiology, Interventional Cardiology
- FDA record
- Classification database entry for DXE
Latest FDA decisions
The 8 newest of 11 since 2015.
K240078 · Decided Aug 28, 2024 · 230 days after receipt
Device type: Catheter, Embolectomy (DXE) · Cardiology
K241330 · Decided Jul 2, 2024 · 53 days after receipt
Device type: Catheter, Embolectomy (DXE) · Cardiology
K233819 · Decided May 22, 2024 · 173 days after receipt
Device type: Catheter, Embolectomy (DXE) · Cardiology
K233820 · Decided May 22, 2024 · 173 days after receipt
Device type: Catheter, Embolectomy (DXE) · Cardiology
K233619 · Decided May 20, 2024 · 189 days after receipt
Device type: Catheter, Embolectomy (DXE) · Cardiology
- Merlin Aspiration System510(k)
K213771 · Decided Jan 27, 2022 · 56 days after receipt
Device type: Catheter, Embolectomy (DXE) · Cardiology
Applied Medical Resources Corporation
K202049 · Decided Sep 7, 2021 · 410 days after receipt
Device type: Catheter, Embolectomy (DXE) · Cardiology
K193379 · Decided Jun 12, 2020 · 190 days after receipt
Device type: Catheter, Embolectomy (DXE) · Cardiology
Reports received
499
Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.
Malfunction reports
478
96% of reports
Injury reports
18
3.6% of reports
Death reports
2
0.4% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 34 | 0 | 0 | 0 |
| Jan 24 | 45 | 1 | 0 | 0 |
| Feb 24 | 41 | 3 | 0 | 0 |
| Mar 24 | 26 | 7 | 1 | 0 |
| Apr 24 | 32 | 0 | 0 | 0 |
| May 24 | 15 | 0 | 0 | 0 |
| Jun 24 | 24 | 2 | 0 | 0 |
| Jul 24 | 31 | 3 | 0 | 0 |
| Aug 24 | 26 | 1 | 0 | 0 |
| Sep 24 | 38 | 2 | 0 | 0 |
| Oct 24 | 16 | 3 | 0 | 0 |
| Nov 24 | 34 | 1 | 0 | 0 |
| Dec 24 | 24 | 1 | 0 | 0 |
| Jan 25 | 32 | 3 | 1 | 0 |
| Feb 25 | 23 | 1 | 0 | 0 |
| Mar 25 | 22 | 2 | 0 | 0 |
| Apr 25 | 24 | 2 | 0 | 0 |
| May 25 | 11 | 1 | 0 | 0 |
| Jun 25 | 12 | 1 | 1 | 0 |
| Jul 25 | 29 | 0 | 0 | 0 |
| Aug 25 | 21 | 2 | 0 | 0 |
| Sep 25 | 36 | 3 | 0 | 0 |
| Oct 25 | 30 | 1 | 0 | 0 |
| Nov 25 | 30 | 4 | 1 | 0 |
| Dec 25 | 15 | 2 | 0 | 0 |
| Jan 26 | 17 | 1 | 0 | 0 |
| Feb 26 | 24 | 1 | 0 | 0 |
| Mar 26 | 41 | 2 | 0 | 0 |
| Apr 26 | 50 | 1 | 0 | 0 |
| May 26 | 29 | 1 | 0 | 0 |
| Jun 26 | 65 | 1 | 0 | 0 |
| Jul 26 | 57 | 2 | 1 | 0 |
| Aug 26 | 52 | 1 | 0 | 0 |
| Sep 26 | 49 | 2 | 0 | 1 |
| Oct 26 | 49 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Material Rupture248
- Material Fragmentation57
- Detachment of Device or Device Component44
- Burst Container or Vessel37
- Adverse Event Without Identified Device or Use Problem21
- Failure to Deflate21
- Material Separation12
- Difficult to Remove11
- Suction Problem11
- Break10
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions474
- Foreign Body In Patient4
- Foreign Body Embolism3
- Renal Failure3
- Thrombosis/Thrombus3
- Chest Pain2
- Insufficient Information2
- Perforation of Vessels2
- Abdominal Pain1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0117-2027 · Started Aug 28, 2026 · Posted by FDA Oct 7, 2026
Reason given: Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
Cause as FDA categorized it: Under Investigation by firm
Z-2526-2026 · Started Apr 21, 2026 · Posted by FDA Jun 23, 2026
Reason given: A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
Cause as FDA categorized it: Under Investigation by firm
Z-1750-2025 · Started Apr 11, 2025 · Posted by FDA May 12, 2025
Reason given: During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
Cause as FDA categorized it: Packaging process control
- Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies.Open, Classified
Z-1824-2024 · Started Apr 17, 2024 · Posted by FDA May 15, 2024
Reason given: The guide tip can become damaged and result in the tip detaching.
Cause as FDA categorized it: Under Investigation by firm
- Fogarty Thru-Lumen Embolectomy Catheter, REF: 12TLW805F35Open, Classified
Z-0238-2024 · Started Sep 7, 2023 · Posted by FDA Nov 6, 2023
Reason given: Inadvertent release of embolectomy catheters that may experience slower than normal balloon deflation (greater-than 15 seconds).
Cause as FDA categorized it: Nonconforming Material/Component
Z-0146-2023 · Started Oct 6, 2022 · Posted by FDA Oct 28, 2022
Reason given: Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches.
Cause as FDA categorized it: Packaging
Z-0147-2023 · Started Oct 6, 2022 · Posted by FDA Oct 28, 2022
Reason given: Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches.
Cause as FDA categorized it: Packaging
Z-0228-2023 · Started Sep 30, 2022 · Posted by FDA Nov 11, 2022
Reason given: Three lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip toward the tip of the catheter and block the inflation hole, leading to the risk of the tip breaking off, and potentially becoming lost in the blood vessel.
Cause as FDA categorized it: Process design
Showing the 8 newest of 58 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.