Device type · DXH
Transmitters And Receivers, Electrocardiograph, Telephone
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
DXH
Cardiovascular
Device class
Class II
FDA decisions since 2015
38
510(k), De Novo and PMA
Adverse event reports
8
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Transmitters And Receivers, Electrocardiograph, Telephone
- Product code
- DXH
- Device class
- Class II
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2920
- Specialty
- Cardiology, Electrophysiology
- FDA record
- Classification database entry for DXH
Latest FDA decisions
The 8 newest of 38 since 2015.
Measure Labs, Inc. (Dba Preemptiveai, Inc.)
K250233 · Decided Feb 13, 2026 · 382 days after receipt
Device type: Transmitters And Receivers, Electrocardiograph, Telephone (DXH) · Cardiology
K250258 · Decided Dec 8, 2025 · 314 days after receipt
Device type: Transmitters And Receivers, Electrocardiograph, Telephone (DXH) · Cardiology
K243438 · Decided Dec 5, 2025 · 394 days after receipt
Device type: Transmitters And Receivers, Electrocardiograph, Telephone (DXH) · Cardiology
K243981 · Decided Nov 18, 2025 · 330 days after receipt
Device type: Transmitters And Receivers, Electrocardiograph, Telephone (DXH) · Cardiology
VitalSigns Technology Co., Ltd.
K243003 · Decided Jun 17, 2025 · 264 days after receipt
Device type: Transmitters And Receivers, Electrocardiograph, Telephone (DXH) · Cardiology
- QT ECG (QTERD100)510(k)
K233521 · Decided Dec 6, 2023 · 35 days after receipt
Device type: Transmitters And Receivers, Electrocardiograph, Telephone (DXH) · Cardiology
- V-Patch Cardiac Monitor510(k)
Shandong Corecare Technology Limited
K222842 · Decided Oct 20, 2022 · 30 days after receipt
Device type: Transmitters And Receivers, Electrocardiograph, Telephone (DXH) · Cardiology
- QT ECG510(k)
K220795 · Decided Sep 26, 2022 · 192 days after receipt
Device type: Transmitters And Receivers, Electrocardiograph, Telephone (DXH) · Cardiology
Reports received
8
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
7
88% of reports
Injury reports
1
13% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 0 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 1 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 0 | 1 | 0 | 0 |
| Jan 25 | 0 | 0 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 12 | 2 | 0 | 0 |
| Sep 25 | 2 | 1 | 0 | 0 |
| Oct 25 | 2 | 1 | 0 | 0 |
| Nov 25 | 0 | 1 | 0 | 0 |
| Dec 25 | 6 | 0 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 1 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 0 | 0 | 0 | 0 |
| Jul 26 | 0 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Connection Problem2
- Electrical /Electronic Property Problem2
- Communication or Transmission Problem1
- Off-Label Use1
- Power Problem1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions7
- Myocardial Infarction1
- Stroke/CVA1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. ElectrocardiographOpen, Classified
GE Medical Systems China Co., Ltd.
Z-2306-2026 · Started Apr 21, 2026 · Posted by FDA Jun 3, 2026
Reason given: Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
Cause as FDA categorized it: Software Manufacturing/Software Deployment
Z-0172-2016 · Started Sep 3, 2015 · Posted by FDA Oct 30, 2015
Reason given: When the battery is inserted into the CardioCall ECG Event Recorder, the device beeps twice and is followed by a continuous ringing sound that only stops when battery is removed. The device has a self-test failure and will not operate normally.
Cause as FDA categorized it: Vendor change control
Z-1125-2015 · Started Jan 9, 2015 · Posted by FDA Feb 18, 2015
Reason given: Alive ECG App version 2.1.2 (intended to be used with the AliveCor Heart Monitor) crashed upon use of the application.
Cause as FDA categorized it: Software change control
Z-0206-2012 · Started Sep 26, 2011 · Posted by FDA Nov 17, 2011
Reason given: LifeWatch received complaints of minor shocks and burns during use of the LifeStar ACT systems with shorts in the leads.
Cause as FDA categorized it: Nonconforming Material/Component
Z-0207-2012 · Started Sep 26, 2011 · Posted by FDA Nov 17, 2011
Reason given: LifeWatch received complaints of minor shocks and burns during use of the LifeStar ACT systems with shorts in the leads.
Cause as FDA categorized it: Nonconforming Material/Component
Z-0208-2012 · Started Sep 26, 2011 · Posted by FDA Nov 17, 2011
Reason given: LifeWatch received complaints of minor shocks and burns during use of the LifeStar ACT systems with shorts in the leads.
Cause as FDA categorized it: Nonconforming Material/Component
Z-0852-2010 · Started Jan 23, 2008 · Posted by FDA Feb 25, 2010
Reason given: A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use.
Cause as FDA categorized it: Component design/selection
Z-0853-2010 · Started Jan 23, 2008 · Posted by FDA Feb 25, 2010
Reason given: A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use.
Cause as FDA categorized it: Component design/selection
Showing the 8 newest of 10 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.