Device type · FET
Endoscopic Video Imaging System/Component, Gastroenterology-Urology
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
FET
Gastroenterology & Urology
Device class
Class II
FDA decisions since 2015
32
510(k), De Novo and PMA
Adverse event reports
17,459
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Endoscopic Video Imaging System/Component, Gastroenterology-Urology
- Product code
- FET
- Device class
- Class II
- Review panel
- Gastroenterology & Urology
- Regulation
- 21 CFR 876.1500
- Specialty
- GI, Urology & Nephrology, Urology
- FDA record
- Classification database entry for FET
Latest FDA decisions
The 8 newest of 32 since 2015.
K262079 · Decided Sep 10, 2026 · 80 days after receipt
Device type: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FET) · GI, Urology & Nephrology
K261880 · Decided Jul 2, 2026 · 27 days after receipt
Device type: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FET) · GI, Urology & Nephrology
Olympus Medical Systems Corporation
K261260 · Decided Jun 24, 2026 · 69 days after receipt
Device type: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FET) · GI, Urology & Nephrology
K251734 · Decided Sep 24, 2025 · 110 days after receipt
Device type: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FET) · GI, Urology & Nephrology
Univlabs Technologies Private Limited
K250462 · Decided Aug 11, 2025 · 174 days after receipt
Device type: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FET) · GI, Urology & Nephrology
- Endoscopic Camera System510(k)
Shenzhen Sophway Technology Co., Ltd.
K250204 · Decided Jul 1, 2025 · 158 days after receipt
Device type: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FET) · GI, Urology & Nephrology
K243260 · Decided Apr 11, 2025 · 178 days after receipt
Device type: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FET) · GI, Urology & Nephrology
Olympus Corporations of the Americas
K242067 · Decided Apr 2, 2025 · 261 days after receipt
Device type: Endoscopic Video Imaging System/Component, Gastroenterology-Urology (FET) · GI, Urology & Nephrology
Reports received
17,459
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
17,393
100% of reports
Injury reports
64
0.4% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 736 | 3 | 0 | 0 |
| Jan 24 | 845 | 9 | 0 | 0 |
| Feb 24 | 1,001 | 9 | 0 | 0 |
| Mar 24 | 1,073 | 2 | 0 | 0 |
| Apr 24 | 1,146 | 5 | 0 | 0 |
| May 24 | 1,292 | 4 | 0 | 0 |
| Jun 24 | 1,170 | 11 | 0 | 0 |
| Jul 24 | 1,379 | 16 | 0 | 0 |
| Aug 24 | 1,184 | 10 | 0 | 0 |
| Sep 24 | 1,072 | 6 | 0 | 0 |
| Oct 24 | 1,254 | 7 | 0 | 0 |
| Nov 24 | 634 | 9 | 0 | 0 |
| Dec 24 | 1,005 | 10 | 0 | 0 |
| Jan 25 | 1,202 | 16 | 0 | 0 |
| Feb 25 | 1,420 | 14 | 0 | 0 |
| Mar 25 | 1,178 | 16 | 0 | 0 |
| Apr 25 | 1,170 | 3 | 0 | 0 |
| May 25 | 826 | 2 | 0 | 0 |
| Jun 25 | 885 | 2 | 0 | 0 |
| Jul 25 | 1,180 | 2 | 0 | 0 |
| Aug 25 | 1,378 | 2 | 0 | 0 |
| Sep 25 | 1,145 | 6 | 0 | 0 |
| Oct 25 | 1,657 | 2 | 0 | 0 |
| Nov 25 | 1,235 | 5 | 0 | 0 |
| Dec 25 | 1,325 | 11 | 0 | 0 |
| Jan 26 | 1,087 | 13 | 0 | 0 |
| Feb 26 | 1,392 | 4 | 0 | 0 |
| Mar 26 | 1,559 | 7 | 0 | 0 |
| Apr 26 | 1,919 | 6 | 0 | 0 |
| May 26 | 1,829 | 4 | 0 | 0 |
| Jun 26 | 1,325 | 2 | 0 | 0 |
| Jul 26 | 1,454 | 3 | 0 | 0 |
| Aug 26 | 1,552 | 2 | 0 | 0 |
| Sep 26 | 1,509 | 3 | 0 | 0 |
| Oct 26 | 1,207 | 4 | 0 | 2 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Shirakawa Olympus17,008
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Display/Image5,032
- Poor Quality Image3,906
- Communication or Transmission Problem3,033
- Erratic or Intermittent Display1,235
- Leak/Splash1,092
- Gas/Air Leak801
- Output Problem672
- Corroded593
- Connection Problem564
- Loose or Intermittent Connection543
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions17,386
- Insufficient Information56
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available10
- Bowel Perforation4
- Burn(s)4
- Hemorrhage/Blood Loss/Bleeding2
- Laceration(s)2
- Superficial (First Degree) Burn2
- Unspecified Tissue Injury2
- Asystole1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Olympus Corporation of the Americas
Z-0181-2025 · Started Aug 21, 2024 · Posted by FDA Oct 23, 2024
Reason given: It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
Cause as FDA categorized it: Under Investigation by firm
Z-0329-2024 · Started Sep 25, 2023 · Posted by FDA Nov 17, 2023
Reason given: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Cause as FDA categorized it: Under Investigation by firm
Z-0346-2024 · Started Sep 25, 2023 · Posted by FDA Nov 17, 2023
Reason given: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Cause as FDA categorized it: Under Investigation by firm
Olympus Corporation of the Americas
Z-2269-2018 · Started May 11, 2018
Reason given: Issuance of validated, new reprocessing procedures. This corrective action is being undertaken as a result of ongoing manufacturer and FDA activities relating to reported patient infections associated with duodenoscopes. The new cleaning procedure requires additional recess flushing and forceps elevator...
Cause as FDA categorized it: Under Investigation by firm
Olympus Corporation of the Americas
Z-2270-2018 · Started May 11, 2018
Reason given: Issuance of validated, new reprocessing procedures. The new cleaning procedure requires additional recess flushing and forceps elevator raising/lowering steps during precleaning, manual cleaning and manual disinfection.
Cause as FDA categorized it: Under Investigation by firm
Z-2134-2016 · Started Jun 20, 2016 · Posted by FDA Jul 6, 2016
Reason given: Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG).
Cause as FDA categorized it: Device Design
Fujifilm Medical Systems U.S.A., Inc.
Z-2805-2015 · Started Jun 30, 2015 · Posted by FDA Sep 22, 2015
Reason given: The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard...
Cause as FDA categorized it: Device Design
Fujifilm Medical Systems U.S.A., Inc.
Z-2806-2015 · Started Jun 30, 2015 · Posted by FDA Sep 22, 2015
Reason given: The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard...
Cause as FDA categorized it: Device Design
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.