Device type · GEX
Powered Laser Surgical Instrument
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
GEX
General & Plastic Surgery
Device class
Class II
FDA decisions since 2015
700
510(k), De Novo and PMA
Adverse event reports
2,886
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Powered Laser Surgical Instrument
- Product code
- GEX
- Device class
- Class II
- Review panel
- General & Plastic Surgery
- Regulation
- 21 CFR 878.4810
- FDA record
- Classification database entry for GEX
Latest FDA decisions
The 8 newest of 700 since 2015.
- Single Use Optical Fiber510(k)
K262886 · Decided Nov 5, 2026 · 80 days after receipt
Device type: Powered Laser Surgical Instrument (GEX) · General & Plastic Surgery
K263063 · Decided Oct 30, 2026 · 59 days after receipt
Device type: Powered Laser Surgical Instrument (GEX) · General & Plastic Surgery
Risu Medical Technology Co., Ltd.
K262349 · Decided Oct 27, 2026 · 109 days after receipt
Device type: Powered Laser Surgical Instrument (GEX) · General & Plastic Surgery
Sanhe PERFECTLASER Technology Co., Ltd.
K262627 · Decided Oct 26, 2026 · 89 days after receipt
Device type: Powered Laser Surgical Instrument (GEX) · General & Plastic Surgery
Beijing Sea Heart International Science and Technologyco.Ltd
K262701 · Decided Oct 22, 2026 · 83 days after receipt
Device type: Powered Laser Surgical Instrument (GEX) · General & Plastic Surgery
Sanhe PERFECTLASER Technology Co., Ltd.
K262628 · Decided Oct 20, 2026 · 83 days after receipt
Device type: Powered Laser Surgical Instrument (GEX) · General & Plastic Surgery
Keyue (Hangzhou) Medical Information Technology Co., Ltd.
K261572 · Decided Oct 13, 2026 · 154 days after receipt
Device type: Powered Laser Surgical Instrument (GEX) · General & Plastic Surgery
Wuhan Phomed Technology Co., Ltd.
K262346 · Decided Oct 8, 2026 · 90 days after receipt
Device type: Powered Laser Surgical Instrument (GEX) · General & Plastic Surgery
Reports received
2,886
Received Nov 1, 2025 to Nov 19, 2026. Includes 16 of another or unstated type.
Malfunction reports
2,331
81% of reports
Injury reports
536
19% of reports
Death reports
3
0.1% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 171 | 35 | 1 | 0 |
| Jan 24 | 124 | 28 | 0 | 2 |
| Feb 24 | 143 | 23 | 1 | 1 |
| Mar 24 | 149 | 28 | 1 | 0 |
| Apr 24 | 135 | 36 | 0 | 0 |
| May 24 | 153 | 40 | 0 | 2 |
| Jun 24 | 172 | 71 | 0 | 1 |
| Jul 24 | 144 | 37 | 0 | 0 |
| Aug 24 | 166 | 36 | 0 | 1 |
| Sep 24 | 151 | 19 | 1 | 0 |
| Oct 24 | 146 | 30 | 1 | 0 |
| Nov 24 | 176 | 26 | 0 | 1 |
| Dec 24 | 161 | 46 | 0 | 1 |
| Jan 25 | 139 | 28 | 0 | 0 |
| Feb 25 | 202 | 34 | 0 | 0 |
| Mar 25 | 185 | 48 | 2 | 0 |
| Apr 25 | 196 | 37 | 0 | 1 |
| May 25 | 177 | 68 | 0 | 0 |
| Jun 25 | 176 | 75 | 1 | 1 |
| Jul 25 | 195 | 70 | 2 | 1 |
| Aug 25 | 217 | 37 | 1 | 0 |
| Sep 25 | 202 | 28 | 0 | 0 |
| Oct 25 | 201 | 77 | 1 | 1 |
| Nov 25 | 177 | 37 | 0 | 0 |
| Dec 25 | 175 | 44 | 0 | 0 |
| Jan 26 | 168 | 31 | 0 | 1 |
| Feb 26 | 192 | 38 | 0 | 0 |
| Mar 26 | 172 | 68 | 0 | 0 |
| Apr 26 | 206 | 61 | 0 | 2 |
| May 26 | 225 | 115 | 0 | 0 |
| Jun 26 | 219 | 36 | 1 | 0 |
| Jul 26 | 211 | 39 | 2 | 0 |
| Aug 26 | 168 | 25 | 0 | 3 |
| Sep 26 | 212 | 14 | 0 | 4 |
| Oct 26 | 206 | 28 | 0 | 6 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Lumenis1,200
- Gyrus Acmi298
- Wavelight254
- Iridex47
- Lutronic45
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Break675
- Nonstandard Device466
- Adverse Event Without Identified Device or Use Problem281
- Output below Specifications232
- Unexpected Therapeutic Results169
- Material Integrity Problem166
- Firing Problem135
- Communication or Transmission Problem120
- Power Problem93
- Failure to Align90
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions2,158
- Insufficient Information170
- Burn(s)126
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available83
- Eye Injury57
- Keratitis48
- Hemorrhage/Blood Loss/Bleeding44
- Foreign Body In Patient27
- Urinary Retention26
- Urinary Incontinence24
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-3110-2026 · Started Aug 6, 2026 · Posted by FDA Sep 4, 2026
Reason given: Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
Cause as FDA categorized it: Device Design
Z-2554-2026 · Started May 5, 2026 · Posted by FDA Jun 24, 2026
Reason given: Software anomaly that causes a false display of error code 5018.
Cause as FDA categorized it: Device Design
Olympus Corporation of the Americas
Z-1694-2026 · Started Feb 27, 2026 · Posted by FDA Mar 30, 2026
Reason given: Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the...
Cause as FDA categorized it: Process control
Olympus Corporation of the Americas
Z-1695-2026 · Started Feb 27, 2026 · Posted by FDA Mar 30, 2026
Reason given: Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the...
Cause as FDA categorized it: Process control
- CLARITY II Laser System; Model No. 1110200210.Open, Classified
Z-1558-2026 · Started Jan 27, 2026 · Posted by FDA Mar 12, 2026
Reason given: Reports of devices sparking/popping and potentially burning patients.
Cause as FDA categorized it: Device Design
Z-1130-2026 · Started Dec 15, 2025 · Posted by FDA Jan 16, 2026
Reason given: The console label for affected devices is missing the "DANGER" symbol.
Cause as FDA categorized it: Error in labeling
Z-2344-2025 · Started Jul 22, 2025 · Posted by FDA Aug 21, 2025
Reason given: The potential for unsterilized product within finished product labeled as sterile.
Cause as FDA categorized it: Nonconforming Material/Component
Olympus Corporation of the Americas
Z-1920-2025 · Started May 6, 2025 · Posted by FDA Jun 6, 2025
Reason given: Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
Cause as FDA categorized it: Under Investigation by firm
Showing the 8 newest of 201 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.