Device type · GXD
Generator, Lesion, Radiofrequency
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
GXD
Neurology
Device class
Class II
FDA decisions since 2015
12
510(k), De Novo and PMA
Adverse event reports
62
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Generator, Lesion, Radiofrequency
- Product code
- GXD
- Device class
- Class II
- Review panel
- Neurology
- Regulation
- 21 CFR 882.4400
- Specialty
- Neurology & Neurosurgery
- FDA record
- Classification database entry for GXD
Latest FDA decisions
The 8 newest of 12 since 2015.
K260337 · Decided Oct 29, 2026 · 269 days after receipt
Device type: Generator, Lesion, Radiofrequency (GXD) · Neurology & Neurosurgery
Boston Scientific Neuromodulation Corporation
K251247 · Decided Aug 7, 2025 · 106 days after receipt
Device type: Generator, Lesion, Radiofrequency (GXD) · Neurology & Neurosurgery
K242057 · Decided Aug 14, 2024 · 30 days after receipt
Device type: Generator, Lesion, Radiofrequency (GXD) · Neurology & Neurosurgery
K232632 · Decided May 24, 2024 · 268 days after receipt
Device type: Generator, Lesion, Radiofrequency (GXD) · Neurology & Neurosurgery
- OneRF Ablation System510(k)
Neuroone Medical Technologies Corp.
K231675 · Decided Dec 6, 2023 · 181 days after receipt
Device type: Generator, Lesion, Radiofrequency (GXD) · Neurology & Neurosurgery
- APEX 6510(k)
K220122 · Decided Mar 15, 2023 · 421 days after receipt
Device type: Generator, Lesion, Radiofrequency (GXD) · Neurology & Neurosurgery
K203293 · Decided Jan 21, 2022 · 438 days after receipt
Device type: Generator, Lesion, Radiofrequency (GXD) · Neurology & Neurosurgery
- IonicRF Generator510(k)
K201610 · Decided Oct 21, 2020 · 128 days after receipt
Device type: Generator, Lesion, Radiofrequency (GXD) · Neurology & Neurosurgery
Reports received
62
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
28
45% of reports
Injury reports
31
50% of reports
Death reports
1
1.6% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 1 | 0 | 0 | 0 |
| Jan 24 | 2 | 2 | 0 | 0 |
| Feb 24 | 1 | 2 | 0 | 0 |
| Mar 24 | 0 | 4 | 0 | 0 |
| Apr 24 | 2 | 5 | 0 | 0 |
| May 24 | 3 | 3 | 0 | 0 |
| Jun 24 | 2 | 0 | 0 | 0 |
| Jul 24 | 2 | 2 | 0 | 0 |
| Aug 24 | 5 | 0 | 0 | 0 |
| Sep 24 | 2 | 1 | 0 | 0 |
| Oct 24 | 14 | 1 | 0 | 0 |
| Nov 24 | 2 | 3 | 0 | 0 |
| Dec 24 | 5 | 0 | 0 | 0 |
| Jan 25 | 9 | 2 | 0 | 0 |
| Feb 25 | 4 | 1 | 0 | 0 |
| Mar 25 | 6 | 4 | 0 | 0 |
| Apr 25 | 7 | 0 | 0 | 0 |
| May 25 | 8 | 3 | 0 | 0 |
| Jun 25 | 2 | 3 | 0 | 0 |
| Jul 25 | 10 | 2 | 0 | 0 |
| Aug 25 | 4 | 1 | 0 | 0 |
| Sep 25 | 3 | 3 | 0 | 0 |
| Oct 25 | 0 | 6 | 0 | 0 |
| Nov 25 | 4 | 2 | 0 | 0 |
| Dec 25 | 2 | 2 | 0 | 0 |
| Jan 26 | 0 | 2 | 0 | 0 |
| Feb 26 | 0 | 2 | 0 | 0 |
| Mar 26 | 1 | 1 | 0 | 0 |
| Apr 26 | 2 | 3 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 10 | 5 | 1 | 1 |
| Jul 26 | 2 | 3 | 0 | 0 |
| Aug 26 | 6 | 6 | 0 | 1 |
| Sep 26 | 1 | 3 | 0 | 0 |
| Oct 26 | 0 | 2 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Connection Problem13
- Therapeutic or Diagnostic Output Failure10
- Insufficient Heating8
- Grounding Malfunction6
- Fracture4
- Impedance Problem4
- Nonstandard Device4
- Adverse Event Without Identified Device or Use Problem3
- Temperature Problem3
- Use of Device Problem3
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions46
- Arrhythmia3
- Burn(s)2
- Electric Shock2
- Asystole1
- Burning Sensation1
- Cardiac Arrest1
- Discomfort1
- Emotional Changes1
- Full thickness (Third Degree) Burn1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0209-2024 · Started Sep 28, 2023 · Posted by FDA Nov 2, 2023
Reason given: Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.
Cause as FDA categorized it: Process change control
Z-2622-2023 · Started Aug 1, 2023 · Posted by FDA Sep 21, 2023
Reason given: Avanos has received an increase in complaints from customers for COOLIEF RF Generators producing F100 and F101 faults.
Cause as FDA categorized it: Process change control
Abbott Laboratories Inc. (St Jude Medical)
Z-1549-2021 · Started Apr 22, 2021
Reason given: Potential capacitor failure which typically manifests as in inability to provide output on one or more channels. In some instances, users reported a charring/smoky odor or emission.
Cause as FDA categorized it: Process change control
Z-0664-2020 · Started Oct 14, 2019
Reason given: Affected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere with patient skin contact and lead to uneven heating.
Cause as FDA categorized it: Process control
Stryker Instruments Div. of Stryker Corporation
Z-0584-2017 · Started Oct 11, 2016 · Posted by FDA Nov 16, 2016
Reason given: Stryker Instruments is voluntarily recalling the Care Instructions/Instructions for Use packaged with the Radiofrequency (RF) MultiGen Generator, because the instructions for use incorrectly contained instructions supporting the sterilization of the MultiGen cables.
Cause as FDA categorized it: Labeling Change Control
Z-1361-2016 · Started Oct 9, 2013 · Posted by FDA Apr 9, 2016
Reason given: NT2000iX software shipped with the international setting turned on. The international settings include access to Corodotomy, Bi-Polar, and No Temperature modes were not cleared in the United States for use.
Cause as FDA categorized it: Software Manufacturing/Software Deployment
Stryker Instruments Div. of Stryker Corporation
Z-0915-2011 · Started Apr 13, 2010 · Posted by FDA Jan 18, 2011
Reason given: This recall is related to the Electri-cord recall. There is a risk when excessive force is applied to the plugs, a fracture can develop on the prongs inside the molded section of the plug. Over time the fracture may lead to arcing inside of the molded section of the plug and a build up of heat which can cause...
Cause as FDA categorized it: Component design/selection
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.