Device type · HQL
Intraocular Lens
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
HQL
Ophthalmic
Device class
Class III
FDA decisions since 2015
4
510(k), De Novo and PMA
Adverse event reports
5,238
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Intraocular Lens
- Product code
- HQL
- Device class
- Class III
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.3600
- Specialty
- Ophthalmology
- FDA record
- Classification database entry for HQL
Latest FDA decisions
- TECNIS PureSee IOL, Model DEN00V TECNIS PureSee Toric II IOLs, Model DET150, DET225, DET300, DET375PMA supplement
Johnson & Johnson Surgical Vision, Inc.
P980040-S176 · Decided Mar 11, 2026 · 223 days after receipt · Panel Track
Device type: intraocular lens (HQL) · Ophthalmology
- RayOne EMV ToricPMA supplement
Rayner Intraocular Lenses , Ltd.
P060011-S039 · Decided Oct 9, 2025 · 206 days after receipt · Panel Track
Device type: intraocular lens (HQL) · Ophthalmology
Beaver Visitec International, Inc.
P240038 · Decided Sep 10, 2025 · 315 days after receipt
Device type: intraocular lens (HQL) · Ophthalmology
P190018 · Decided Jan 7, 2020 · 180 days after receipt
Device type: intraocular lens (HQL) · Ophthalmology
Reports received
5,238
Received Nov 1, 2025 to Nov 19, 2026. Includes 9 of another or unstated type.
Malfunction reports
4,076
78% of reports
Injury reports
1,153
22% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 278 | 101 | 0 | 1 |
| Jan 24 | 170 | 82 | 0 | 0 |
| Feb 24 | 247 | 84 | 0 | 0 |
| Mar 24 | 232 | 131 | 0 | 0 |
| Apr 24 | 220 | 96 | 0 | 0 |
| May 24 | 249 | 101 | 0 | 0 |
| Jun 24 | 221 | 76 | 0 | 1 |
| Jul 24 | 256 | 114 | 0 | 0 |
| Aug 24 | 215 | 92 | 0 | 0 |
| Sep 24 | 240 | 93 | 0 | 0 |
| Oct 24 | 287 | 72 | 0 | 0 |
| Nov 24 | 271 | 80 | 0 | 0 |
| Dec 24 | 419 | 113 | 0 | 0 |
| Jan 25 | 299 | 54 | 0 | 0 |
| Feb 25 | 393 | 65 | 0 | 0 |
| Mar 25 | 367 | 78 | 0 | 0 |
| Apr 25 | 441 | 131 | 0 | 0 |
| May 25 | 375 | 105 | 0 | 1 |
| Jun 25 | 312 | 70 | 0 | 0 |
| Jul 25 | 413 | 103 | 0 | 0 |
| Aug 25 | 353 | 78 | 0 | 0 |
| Sep 25 | 298 | 86 | 0 | 0 |
| Oct 25 | 321 | 79 | 0 | 1 |
| Nov 25 | 361 | 93 | 0 | 0 |
| Dec 25 | 382 | 129 | 0 | 0 |
| Jan 26 | 306 | 64 | 0 | 0 |
| Feb 26 | 315 | 92 | 0 | 0 |
| Mar 26 | 335 | 73 | 0 | 0 |
| Apr 26 | 425 | 110 | 0 | 0 |
| May 26 | 337 | 92 | 0 | 0 |
| Jun 26 | 350 | 142 | 0 | 0 |
| Jul 26 | 362 | 101 | 0 | 0 |
| Aug 26 | 301 | 103 | 0 | 5 |
| Sep 26 | 320 | 68 | 0 | 3 |
| Oct 26 | 282 | 86 | 0 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Amo Puerto Rico Mfg1,526
- Physiol10
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Break978
- Adverse Event Without Identified Device or Use Problem796
- Device Difficult to Setup or Prepare551
- Scratched Material436
- Material Twisted/Bent353
- Failure to Deliver351
- Material Split, Cut or Torn314
- Failure to Advance284
- Device Damaged by Another Device219
- Crack166
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions2,591
- Insufficient Information1,555
- Visual Impairment307
- Blurred Vision294
- Visual Disturbances212
- Capsular Bag Tear125
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available104
- Eye Injury95
- Inflammation78
- Posterior Capsule Opacification65
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
AMO Puerto Rico Manufacturing, Inc.
Z-1634-2026 · Started Feb 19, 2026 · Posted by FDA Mar 24, 2026
Reason given: Due to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding as consistently as expected.
Cause as FDA categorized it: Process control
- Akreos, SKU: AO60P0300. Akreos intraocular lensesOpen, Classified
Z-2451-2025 · Started Jun 6, 2025 · Posted by FDA Aug 26, 2025
Reason given: An off-power error was identified, which resulted in a diopter of 16, rather than the intended 3 diopter. Implantation of affected intraocular lens could result in functional visual impairment or reduction in visual acuity which could result in the need for optical or surgical correction.
Cause as FDA categorized it: Process control
Z-1679-2025 · Started Mar 27, 2025 · Posted by FDA May 2, 2025
Reason given: In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Cause as FDA categorized it: Process control
- TECNIS Eyhance IOL, REF: ICB00, STERILE.EO, Rx OnlyOpen, Classified
Johnson & Johnson Surgical Vision, Inc.
Z-2289-2023 · Started Jul 6, 2023 · Posted by FDA Aug 3, 2023
Reason given: Their is a potential that intraocular lenses (IOLs) were manufactured with unintended toric fiducial marks.
Cause as FDA categorized it: Under Investigation by firm
Z-1338-2022 · Started May 13, 2022
Reason given: The outer package is mislabeled and the package contains a different IOL strength.
Cause as FDA categorized it: Under Investigation by firm
Alcon Laboratories Ireland, Ltd
Z-1269-2022 · Started May 11, 2022
Reason given: Due to device (preloaded intraocular lens) mislabeling, associated with partial mixing of product during overlapping manufacturing runs, and attributed to improper manufacturing line clearance.
Cause as FDA categorized it: Under Investigation by firm
Z-2552-2020 · Started Jun 9, 2020
Reason given: The toric axis marks were misaligned (from the expected position in relation to the haptics) on an implanted lens.
Cause as FDA categorized it: Unknown/Undetermined by firm
Z-2287-2020 · Started Apr 17, 2020
Reason given: Incorrect IOL diopter
Cause as FDA categorized it: Process control
Showing the 8 newest of 75 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.