Device type · MHX
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
MHX
Cardiovascular
Device class
Class II
FDA decisions since 2015
126
510(k), De Novo and PMA
Adverse event reports
5,029
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Product code
- MHX
- Device class
- Class II
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
- Specialty
- Cardiology, Electrophysiology
- FDA record
- Classification database entry for MHX
Latest FDA decisions
The 8 newest of 126 since 2015.
- 12SL Precision510(k)
Ge Medical Systems Information Technologies, Inc.
K261307 · Decided Nov 3, 2026 · 196 days after receipt
Device type: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) (MHX) · Cardiology
K260676 · Decided Oct 14, 2026 · 226 days after receipt
Device type: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) (MHX) · Cardiology
K260656 · Decided Sep 1, 2026 · 186 days after receipt
Device type: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) (MHX) · Cardiology
K260343 · Decided Aug 28, 2026 · 207 days after receipt
Device type: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) (MHX) · Cardiology
K260579 · Decided Aug 20, 2026 · 181 days after receipt
Device type: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) (MHX) · Cardiology
K251508 · Decided Aug 6, 2026 · 447 days after receipt
Device type: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) (MHX) · Cardiology
Ge Medical Systems Information Technologies, Inc.
K253277 · Decided Jul 24, 2026 · 298 days after receipt
Device type: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) (MHX) · Cardiology
K254209 · Decided Jun 4, 2026 · 157 days after receipt
Device type: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) (MHX) · Cardiology
Reports received
5,029
Received Nov 1, 2025 to Nov 19, 2026. Includes 18 of another or unstated type.
Malfunction reports
3,974
79% of reports
Injury reports
915
18% of reports
Death reports
122
2.4% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 173 | 60 | 12 | 0 |
| Jan 24 | 164 | 42 | 5 | 0 |
| Feb 24 | 164 | 36 | 12 | 2 |
| Mar 24 | 185 | 54 | 11 | 0 |
| Apr 24 | 173 | 52 | 6 | 0 |
| May 24 | 196 | 49 | 3 | 0 |
| Jun 24 | 181 | 64 | 3 | 0 |
| Jul 24 | 169 | 50 | 1 | 0 |
| Aug 24 | 166 | 49 | 3 | 0 |
| Sep 24 | 156 | 31 | 5 | 0 |
| Oct 24 | 203 | 56 | 7 | 0 |
| Nov 24 | 189 | 41 | 8 | 0 |
| Dec 24 | 218 | 50 | 10 | 0 |
| Jan 25 | 180 | 42 | 6 | 0 |
| Feb 25 | 166 | 42 | 9 | 0 |
| Mar 25 | 216 | 78 | 9 | 0 |
| Apr 25 | 231 | 48 | 5 | 0 |
| May 25 | 189 | 59 | 4 | 0 |
| Jun 25 | 211 | 75 | 13 | 0 |
| Jul 25 | 332 | 52 | 9 | 0 |
| Aug 25 | 249 | 57 | 8 | 0 |
| Sep 25 | 262 | 63 | 7 | 0 |
| Oct 25 | 355 | 73 | 10 | 0 |
| Nov 25 | 246 | 75 | 8 | 0 |
| Dec 25 | 344 | 66 | 14 | 0 |
| Jan 26 | 264 | 69 | 6 | 0 |
| Feb 26 | 337 | 52 | 18 | 0 |
| Mar 26 | 360 | 84 | 9 | 0 |
| Apr 26 | 333 | 70 | 12 | 0 |
| May 26 | 337 | 97 | 7 | 0 |
| Jun 26 | 336 | 91 | 12 | 0 |
| Jul 26 | 395 | 83 | 14 | 0 |
| Aug 26 | 299 | 86 | 7 | 14 |
| Sep 26 | 392 | 84 | 11 | 1 |
| Oct 26 | 331 | 58 | 4 | 3 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Philips2,719
- Nihon Kohden575
- Masimo14
- Baxter10
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Audible Alarm845
- Patient Device Interaction Problem770
- Incorrect Measurement487
- Defective Alarm323
- No Audible Prompt/Feedback272
- Unable to Obtain Readings201
- Unintended Application Program Shut Down180
- Application Program Problem172
- Output Problem162
- Device Alarm System149
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions3,992
- Skin Inflammation/ Irritation772
- Insufficient Information177
- Low Oxygen Saturation34
- Cardiac Arrest24
- Tachycardia20
- Asystole17
- Bradycardia17
- Arrhythmia14
- Hypoxia11
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0199-2027 · Started Sep 16, 2026 · Posted by FDA Oct 16, 2026
Reason given: Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).
Cause as FDA categorized it: Software design
Z-0089-2027 · Started Sep 8, 2026 · Posted by FDA Oct 6, 2026
Reason given: Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.
Cause as FDA categorized it: Software design
- ePM Series Patient Monitor. Patient Monitor Series: ePM 10M/ePM 12M/ePM 12MA/ ePM 15M/ePM 10AOpen, Classified
Z-0109-2027 · Started Aug 31, 2026 · Posted by FDA Oct 7, 2026
Reason given: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Cause as FDA categorized it: Labeling design
Z-0110-2027 · Started Aug 31, 2026 · Posted by FDA Oct 7, 2026
Reason given: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Cause as FDA categorized it: Labeling design
- BeneView Series. Operator's Manuals: T5 / T8Open, Classified
Z-0113-2027 · Started Aug 31, 2026 · Posted by FDA Oct 7, 2026
Reason given: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Cause as FDA categorized it: Labeling design
Z-0114-2027 · Started Aug 31, 2026 · Posted by FDA Oct 7, 2026
Reason given: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Cause as FDA categorized it: Labeling design
- cPM Series Patient Monitor. Operator's Manual: cPM 8/cPM 12Open, Classified
Z-0115-2027 · Started Aug 31, 2026 · Posted by FDA Oct 7, 2026
Reason given: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Cause as FDA categorized it: Labeling design
- T1 Patient Monitor. Operator's Manual: T1 Patient MonitorOpen, Classified
Z-0116-2027 · Started Aug 31, 2026 · Posted by FDA Oct 7, 2026
Reason given: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Cause as FDA categorized it: Labeling design
Showing the 8 newest of 294 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.