MedTech Resources

Device type · MSD

Catheter, Hemodialysis, Implanted

FDA groups similar devices under a three-letter product code. This page gathers what FDA's public data says about MSD: how it is classified, what has been cleared or approved, and what has been reported and recalled.

Last checked: Nov 19, 2026Source: openFDAMethods & limitations

Product code

MSD

Gastroenterology & Urology

Device class

Class II

FDA decisions since 2015

17

510(k), De Novo and PMA

Adverse event reports

776

Received Nov 1, 2025 to Nov 19, 2026

How FDA classifies it

Device type
Catheter, Hemodialysis, Implanted
Product code
MSD
Device class
Class II
Review panel
Gastroenterology & Urology
Regulation
21 CFR 876.5540

Latest FDA decisions

The 8 newest of 17 since 2015.

All decisions for MSD

Reports received

776

Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.

Malfunction reports

656

85% of reports

Injury reports

111

14% of reports

Death reports

8

1.0% of reports

Adverse event reports received per month

By the date FDA received each report, stacked by event type. The newest month is incomplete.

Open in the MAUDE Monitor
  • Malfunction
  • Injury
  • Death
  • Other
View as table
MonthMalfunctionInjuryDeathOther
Dec 23451300
Jan 2459330
Feb 2438300
Mar 2426700
Apr 2441500
May 2440400
Jun 2454400
Jul 24721910
Aug 24891200
Sep 24641210
Oct 2476510
Nov 24541800
Dec 24844420
Jan 25933600
Feb 2565820
Mar 25112720
Apr 2587400
May 25105700
Jun 25119410
Jul 25942710
Aug 25511830
Sep 2546810
Oct 2542910
Nov 25591520
Dec 2539210
Jan 2633600
Feb 26126800
Mar 26664710
Apr 261691320
May 2631210
Jun 2633210
Jul 2626200
Aug 2624600
Sep 2629501
Oct 2621300
Nov 260000

Manufacturers named most often

Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.

Device problems as coded

Last 12 months. FDA's terms. A report can carry several.

  • Deformation Due to Compressive Stress155
  • Fracture121
  • Fluid/Blood Leak101
  • Material Deformation81
  • Material Discolored73
  • Expulsion69
  • Loosening of Implant Not Related to Bone-Ingrowth66
  • Material Protrusion/Extrusion49
  • Adverse Event Without Identified Device or Use Problem41
  • Failure to Advance34

Patient problems as coded

Last 12 months. FDA's terms. A report can carry several.

  • No Clinical Signs, Symptoms or Conditions654
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available36
  • Hemorrhage/Blood Loss/Bleeding30
  • Insufficient Information15
  • Unspecified Infection10
  • Thrombosis/Thrombus9
  • Air Embolism7
  • Discomfort6
  • Bacterial Infection4
  • Cardiac Arrest4

Recent recalls

From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.

Showing the 8 newest of 270 recall records in FDA's data.

Back to all tools

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.