Device type · MTD
Catheter, Intracardiac Mapping, High-Density Array
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
MTD
Cardiovascular
Device class
Class II
FDA decisions since 2015
18
510(k), De Novo and PMA
Adverse event reports
346
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Catheter, Intracardiac Mapping, High-Density Array
- Product code
- MTD
- Device class
- Class II
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1220
- Specialty
- Cardiology, Electrophysiology
- FDA record
- Classification database entry for MTD
Latest FDA decisions
The 8 newest of 18 since 2015.
K261012 · Decided Apr 23, 2026 · 27 days after receipt
Device type: Catheter, Intracardiac Mapping, High-Density Array (MTD) · Cardiology
K253205 · Decided Mar 13, 2026 · 168 days after receipt
Device type: Catheter, Intracardiac Mapping, High-Density Array (MTD) · Cardiology
K233433 · Decided Aug 7, 2024 · 300 days after receipt
Device type: Catheter, Intracardiac Mapping, High-Density Array (MTD) · Cardiology
K241372 · Decided Jul 9, 2024 · 55 days after receipt
Device type: Catheter, Intracardiac Mapping, High-Density Array (MTD) · Cardiology
K231312 · Decided Jul 10, 2023 · 66 days after receipt
Device type: Catheter, Intracardiac Mapping, High-Density Array (MTD) · Cardiology
K223666 · Decided Mar 17, 2023 · 100 days after receipt
Device type: Catheter, Intracardiac Mapping, High-Density Array (MTD) · Cardiology
K230253 · Decided Mar 2, 2023 · 30 days after receipt
Device type: Catheter, Intracardiac Mapping, High-Density Array (MTD) · Cardiology
K210766 · Decided Jun 21, 2022 · 463 days after receipt
Device type: Catheter, Intracardiac Mapping, High-Density Array (MTD) · Cardiology
Reports received
346
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
282
82% of reports
Injury reports
60
17% of reports
Death reports
4
1.2% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 19 | 4 | 0 | 0 |
| Jan 24 | 33 | 7 | 0 | 0 |
| Feb 24 | 19 | 7 | 1 | 0 |
| Mar 24 | 30 | 6 | 0 | 0 |
| Apr 24 | 19 | 0 | 0 | 0 |
| May 24 | 22 | 10 | 0 | 0 |
| Jun 24 | 28 | 8 | 1 | 0 |
| Jul 24 | 32 | 4 | 1 | 0 |
| Aug 24 | 34 | 4 | 0 | 0 |
| Sep 24 | 33 | 6 | 0 | 0 |
| Oct 24 | 9 | 10 | 1 | 0 |
| Nov 24 | 13 | 9 | 1 | 0 |
| Dec 24 | 25 | 5 | 0 | 0 |
| Jan 25 | 19 | 3 | 0 | 0 |
| Feb 25 | 16 | 5 | 0 | 0 |
| Mar 25 | 18 | 6 | 0 | 0 |
| Apr 25 | 21 | 7 | 0 | 0 |
| May 25 | 28 | 10 | 0 | 0 |
| Jun 25 | 29 | 5 | 0 | 0 |
| Jul 25 | 34 | 13 | 1 | 0 |
| Aug 25 | 36 | 12 | 1 | 0 |
| Sep 25 | 26 | 8 | 1 | 0 |
| Oct 25 | 25 | 4 | 0 | 0 |
| Nov 25 | 29 | 1 | 0 | 0 |
| Dec 25 | 32 | 3 | 1 | 0 |
| Jan 26 | 26 | 9 | 0 | 0 |
| Feb 26 | 27 | 4 | 1 | 0 |
| Mar 26 | 23 | 7 | 0 | 0 |
| Apr 26 | 24 | 5 | 0 | 0 |
| May 26 | 17 | 5 | 1 | 0 |
| Jun 26 | 22 | 4 | 0 | 0 |
| Jul 26 | 26 | 5 | 1 | 0 |
| Aug 26 | 22 | 5 | 0 | 0 |
| Sep 26 | 14 | 6 | 0 | 0 |
| Oct 26 | 20 | 6 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Abbott28
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Obstruction of Flow165
- Patient Device Interaction Problem54
- Entrapment of Device17
- Adverse Event Without Identified Device or Use Problem14
- Contamination /Decontamination Problem14
- Material Twisted/Bent13
- Detachment of Device or Device Component10
- Manufacturing, Packaging or Shipping Problem9
- Device-Device Incompatibility8
- Signal Artifact/Noise8
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions282
- Cardiac Tamponade22
- Cardiac Perforation10
- Cardiac Arrest9
- Foreign Body In Patient6
- Heart Block4
- Ischemic Heart Disease4
- Air Embolism3
- Unspecified Tissue Injury3
- Bradycardia2
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- FIRMap" Catheter, 60mm BasketTerminated
Z-1213-2022 · Started Apr 28, 2022
Reason given: Incorrect product labeling.
Cause as FDA categorized it: Process control
Z-2371-2020 · Started Apr 22, 2020
Reason given: Mapping Catheter can become entrapped in the cardiac valvular apparatus while the user is attempting to create an electroanatomic map.
Cause as FDA categorized it: Error in labeling
Z-0211-2018 · Started May 17, 2017
Reason given: Contraindication language updated in the Instructions For Use (IFU) to indicate that Pentaray Catheters should not be used in patients with prosthetic valves.
Cause as FDA categorized it: Other
Z-0212-2018 · Started May 17, 2017
Reason given: Contraindication language updated in the Instructions For Use (IFU) to indicate that Pentaray Catheters should not be used in patients with prosthetic valves.
Cause as FDA categorized it: Other
Z-1590-2016 · Started Apr 14, 2016
Reason given: Biosense Webster is recalling the Pentaray Catheters to clarify the contraindication language in the Instructions For Use (IFU) and product labeling for this catheter relative to patients with prosthetic valves.
Cause as FDA categorized it: Use error
Z-1591-2016 · Started Apr 14, 2016
Reason given: Biosense Webster is recalling the Pentaray Catheters to clarify the contraindication language in the Instructions For Use (IFU) and product labeling for this catheter relative to patients with prosthetic valves.
Cause as FDA categorized it: Use error
Z-2478-2015 · Started Aug 12, 2015 · Posted by FDA Aug 28, 2015
Reason given: The expiration date was incorrectly printed on the pouch/box labels of 19 products that were shipped to customer sites. The manufacturing date was accidentally used (copied) for the expiration date.
Cause as FDA categorized it: Incorrect or no expiration date
Z-1176-2014 · Started Feb 20, 2014 · Posted by FDA Mar 11, 2014
Reason given: Biosense Webster is recalling the PentaRay Nav Catheter because it has the potential to have partial tip separation from the catheter shaft. Biosense Webster is expanding the recall on 5/1/14 to include all lots of the PentaRay Nav Catheter and the PentaRay Nav ECO Catheter because they can lead to an insufficient...
Cause as FDA categorized it: Process design
Showing the 8 newest of 9 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.