Device type · NOU
Continuous, Ventilator, Home Use
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
NOU
Anesthesiology
Device class
Class II
FDA decisions since 2015
3
510(k), De Novo and PMA
Adverse event reports
916
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Continuous, Ventilator, Home Use
- Product code
- NOU
- Device class
- Class II
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
- Specialty
- Anesthesiology & Respiratory, Pulmonology
- FDA record
- Classification database entry for NOU
Latest FDA decisions
- Vivo 45 LS510(k)
K233452 · Decided Jan 13, 2025 · 451 days after receipt
Device type: Continuous, Ventilator, Home Use (NOU) · Anesthesiology & Respiratory
- Vivo 45 LS510(k)
K193586 · Decided Jan 27, 2021 · 401 days after receipt
Device type: Continuous, Ventilator, Home Use (NOU) · Anesthesiology & Respiratory
- Vivo 60510(k)
K160481 · Decided Nov 9, 2016 · 261 days after receipt
Device type: Continuous, Ventilator, Home Use (NOU) · Anesthesiology & Respiratory
Reports received
916
Received Nov 1, 2025 to Nov 19, 2026. Includes 5 of another or unstated type.
Malfunction reports
844
92% of reports
Injury reports
42
4.6% of reports
Death reports
25
2.7% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 92 | 6 | 1 | 0 |
| Jan 24 | 38 | 1 | 1 | 0 |
| Feb 24 | 46 | 0 | 2 | 0 |
| Mar 24 | 62 | 2 | 1 | 0 |
| Apr 24 | 42 | 0 | 1 | 0 |
| May 24 | 52 | 6 | 1 | 0 |
| Jun 24 | 82 | 0 | 0 | 0 |
| Jul 24 | 74 | 3 | 2 | 0 |
| Aug 24 | 98 | 0 | 5 | 0 |
| Sep 24 | 62 | 1 | 3 | 0 |
| Oct 24 | 80 | 7 | 0 | 0 |
| Nov 24 | 80 | 4 | 2 | 0 |
| Dec 24 | 55 | 3 | 1 | 0 |
| Jan 25 | 61 | 1 | 0 | 0 |
| Feb 25 | 44 | 0 | 0 | 0 |
| Mar 25 | 85 | 4 | 0 | 0 |
| Apr 25 | 42 | 8 | 0 | 0 |
| May 25 | 54 | 2 | 0 | 0 |
| Jun 25 | 35 | 12 | 0 | 0 |
| Jul 25 | 93 | 17 | 0 | 0 |
| Aug 25 | 94 | 2 | 0 | 0 |
| Sep 25 | 67 | 5 | 4 | 0 |
| Oct 25 | 168 | 5 | 7 | 0 |
| Nov 25 | 71 | 1 | 1 | 0 |
| Dec 25 | 46 | 5 | 1 | 0 |
| Jan 26 | 58 | 4 | 4 | 0 |
| Feb 26 | 61 | 4 | 1 | 0 |
| Mar 26 | 70 | 5 | 1 | 0 |
| Apr 26 | 73 | 4 | 0 | 0 |
| May 26 | 42 | 4 | 1 | 0 |
| Jun 26 | 78 | 0 | 3 | 0 |
| Jul 26 | 83 | 2 | 1 | 0 |
| Aug 26 | 87 | 4 | 6 | 1 |
| Sep 26 | 122 | 5 | 3 | 2 |
| Oct 26 | 53 | 4 | 3 | 2 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- ResMed870
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Protective Measures Problem515
- Inadequate User Interface107
- Battery Problem83
- Complete Loss of Power59
- Power Problem52
- Failure to Power Up44
- Charging Problem40
- Unexpected Shutdown30
- Device Alarm System23
- Failure to Charge16
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions862
- Insufficient Information16
- Low Oxygen Saturation13
- Cyanosis9
- Dyspnea5
- Loss of consciousness5
- Cardiac Arrest4
- Respiratory Failure4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
- Diaphoresis2
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1942-2026 · Started Mar 25, 2026 · Posted by FDA May 1, 2026
Reason given: Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to...
Cause as FDA categorized it: Process change control
Z-2690-2024 · Started Jul 25, 2024 · Posted by FDA Sep 3, 2024
Reason given: There is a potential for short term (<7 days) elevated levels of formaldehyde emitted from the device into the patient breathing airpath in specific conditions.
Cause as FDA categorized it: Under Investigation by firm
Z-2942-2024 · Started Jul 16, 2024 · Posted by FDA Sep 5, 2024
Reason given: Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.
Cause as FDA categorized it: Device Design
Z-2944-2024 · Started Jul 16, 2024 · Posted by FDA Sep 5, 2024
Reason given: Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.
Cause as FDA categorized it: Device Design
Z-0405-2024 · Started Oct 13, 2023 · Posted by FDA Nov 27, 2023
Reason given: Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
Cause as FDA categorized it: Process control
Z-1382-2022 · Started Jun 2, 2022
Reason given: A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained
Cause as FDA categorized it: Device Design
Z-2287-2021 · Started Jul 26, 2021
Reason given: Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
Cause as FDA categorized it: Labeling design
Z-2288-2021 · Started Jul 26, 2021
Reason given: Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
Cause as FDA categorized it: Labeling design
Showing the 8 newest of 16 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.