Device type · NRY
Catheter, Thrombus Retriever
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
NRY
Neurology
Device class
Class II
FDA decisions since 2015
95
510(k), De Novo and PMA
Adverse event reports
466
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Catheter, Thrombus Retriever
- Product code
- NRY
- Device class
- Class II
- Review panel
- Neurology
- Regulation
- 21 CFR 870.1250
- Specialty
- Neurology & Neurosurgery
- FDA record
- Classification database entry for NRY
Latest FDA decisions
The 8 newest of 95 since 2015.
K262021 · Decided Oct 21, 2026 · 127 days after receipt
Device type: Catheter, Thrombus Retriever (NRY) · Cardiology
K263266 · Decided Oct 16, 2026 · 28 days after receipt
Device type: Catheter, Thrombus Retriever (NRY) · Cardiology
- Radical 064 Reperfusion Catheter; Radical 076 Reperfusion Catheter; Radical Aspiration Tubing Kit510(k)
K261089 · Decided Sep 14, 2026 · 166 days after receipt
Device type: Catheter, Thrombus Retriever (NRY) · Cardiology
K260617 · Decided Aug 28, 2026 · 184 days after receipt
Device type: Catheter, Thrombus Retriever (NRY) · Cardiology
K262398 · Decided Aug 13, 2026 · 30 days after receipt
Device type: Catheter, Thrombus Retriever (NRY) · Cardiology
K262391 · Decided Aug 12, 2026 · 29 days after receipt
Device type: Catheter, Thrombus Retriever (NRY) · Cardiology
- Millipede System510(k)
K262399 · Decided Aug 12, 2026 · 29 days after receipt
Device type: Catheter, Thrombus Retriever (NRY) · Cardiology
K260002 · Decided Jul 16, 2026 · 195 days after receipt
Device type: Catheter, Thrombus Retriever (NRY) · Cardiology
Reports received
466
Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.
Malfunction reports
326
70% of reports
Injury reports
106
23% of reports
Death reports
33
7.1% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 17 | 43 | 11 | 0 |
| Jan 24 | 24 | 39 | 17 | 0 |
| Feb 24 | 21 | 22 | 4 | 0 |
| Mar 24 | 25 | 25 | 6 | 0 |
| Apr 24 | 27 | 24 | 11 | 0 |
| May 24 | 16 | 30 | 3 | 0 |
| Jun 24 | 17 | 17 | 2 | 0 |
| Jul 24 | 23 | 36 | 7 | 0 |
| Aug 24 | 12 | 28 | 6 | 0 |
| Sep 24 | 13 | 13 | 3 | 0 |
| Oct 24 | 16 | 29 | 6 | 0 |
| Nov 24 | 18 | 18 | 0 | 0 |
| Dec 24 | 58 | 17 | 8 | 0 |
| Jan 25 | 40 | 19 | 9 | 0 |
| Feb 25 | 32 | 11 | 1 | 0 |
| Mar 25 | 37 | 5 | 2 | 0 |
| Apr 25 | 52 | 17 | 3 | 0 |
| May 25 | 30 | 24 | 0 | 0 |
| Jun 25 | 36 | 11 | 2 | 0 |
| Jul 25 | 35 | 17 | 5 | 0 |
| Aug 25 | 25 | 14 | 3 | 0 |
| Sep 25 | 25 | 8 | 0 | 0 |
| Oct 25 | 19 | 7 | 2 | 0 |
| Nov 25 | 30 | 12 | 5 | 0 |
| Dec 25 | 18 | 13 | 3 | 0 |
| Jan 26 | 23 | 6 | 1 | 0 |
| Feb 26 | 27 | 5 | 6 | 0 |
| Mar 26 | 31 | 11 | 4 | 0 |
| Apr 26 | 38 | 10 | 2 | 0 |
| May 26 | 32 | 9 | 1 | 0 |
| Jun 26 | 24 | 10 | 2 | 0 |
| Jul 26 | 28 | 4 | 2 | 0 |
| Aug 26 | 29 | 12 | 5 | 0 |
| Sep 26 | 23 | 10 | 2 | 1 |
| Oct 26 | 23 | 4 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Fracture109
- Adverse Event Without Identified Device or Use Problem96
- Break78
- Physical Resistance/Sticking75
- Material Deformation52
- Retraction Problem47
- Deformation Due to Compressive Stress45
- Material Separation32
- Detachment of Device or Device Component14
- Failure to Advance14
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions330
- Intracranial Hemorrhage43
- Ischemia Stroke28
- Foreign Body In Patient18
- Perforation of Vessels17
- Thromboembolism17
- Vascular Dissection16
- Insufficient Information15
- Hemorrhage/Blood Loss/Bleeding13
- Obstruction/Occlusion13
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-3331-2026 · Started Jul 24, 2026 · Posted by FDA Sep 29, 2026
Reason given: Due to manufacturing issues reperfusion catheters may fracture.
Cause as FDA categorized it: Under Investigation by firm
- Penumbra System RED 68 Reperfusion Catheter REF: RED68KITOpen, Classified
Z-2770-2026 · Started Jun 3, 2026 · Posted by FDA Jul 21, 2026
Reason given: Due to manufacturing issues, reperfusion catheters may fractures
Cause as FDA categorized it: Under Investigation by firm
Z-1512-2025 · Started Feb 26, 2025 · Posted by FDA Apr 7, 2025
Reason given: Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur.
Cause as FDA categorized it: Device Design
Z-1234-2025 · Started Jan 2, 2025 · Posted by FDA Feb 26, 2025
Reason given: Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
Cause as FDA categorized it: Under Investigation by firm
Z-3157-2024 · Started Aug 5, 2024 · Posted by FDA Sep 17, 2024
Reason given: Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date.
Cause as FDA categorized it: Under Investigation by firm
- Trevo ProVue, Catalog: 90184Open, Classified
Z-1903-2024 · Started Apr 23, 2024 · Posted by FDA May 28, 2024
Reason given: Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and...
Cause as FDA categorized it: Process design
Z-1611-2024 · Started Mar 8, 2024 · Posted by FDA Apr 24, 2024
Reason given: Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
Cause as FDA categorized it: Process control
Z-2577-2021 · Started Aug 13, 2021 · Posted by FDA Sep 29, 2021
Reason given: There is a potential for distal end of catheters to fracture and become detached.
Cause as FDA categorized it: Under Investigation by firm
Showing the 8 newest of 27 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.