Device type · OLO
Orthopedic Stereotaxic Instrument
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
OLO
Orthopedic
Device class
Class II
FDA decisions since 2015
390
510(k), De Novo and PMA
Adverse event reports
2,954
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Orthopedic Stereotaxic Instrument
- Product code
- OLO
- Device class
- Class II
- Review panel
- Orthopedic
- Regulation
- 21 CFR 882.4560
- FDA record
- Classification database entry for OLO
Latest FDA decisions
The 8 newest of 390 since 2015.
- GPS Ankle510(k)
K262567 · Decided Oct 26, 2026 · 94 days after receipt
Device type: Orthopedic Stereotaxic Instrument (OLO) · Neurology & Neurosurgery
K260403 · Decided Oct 19, 2026 · 255 days after receipt
Device type: Orthopedic Stereotaxic Instrument (OLO) · Neurology & Neurosurgery
K263234 · Decided Oct 16, 2026 · 30 days after receipt
Device type: Orthopedic Stereotaxic Instrument (OLO) · Neurology & Neurosurgery
K263154 · Decided Oct 9, 2026 · 30 days after receipt
Device type: Orthopedic Stereotaxic Instrument (OLO) · Neurology & Neurosurgery
K262257 · Decided Sep 30, 2026 · 90 days after receipt
Device type: Orthopedic Stereotaxic Instrument (OLO) · Neurology & Neurosurgery
K262506 · Decided Aug 20, 2026 · 30 days after receipt
Device type: Orthopedic Stereotaxic Instrument (OLO) · Neurology & Neurosurgery
Suite 201, National Innovation Centre
K261410 · Decided Jul 30, 2026 · 91 days after receipt
Device type: Orthopedic Stereotaxic Instrument (OLO) · Neurology & Neurosurgery
K261205 · Decided Jul 29, 2026 · 107 days after receipt
Device type: Orthopedic Stereotaxic Instrument (OLO) · Neurology & Neurosurgery
Reports received
2,954
Received Nov 1, 2025 to Nov 19, 2026. Includes 10 of another or unstated type.
Malfunction reports
2,651
90% of reports
Injury reports
289
9.8% of reports
Death reports
4
0.1% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 168 | 23 | 0 | 0 |
| Jan 24 | 120 | 16 | 0 | 0 |
| Feb 24 | 161 | 35 | 0 | 0 |
| Mar 24 | 198 | 21 | 1 | 0 |
| Apr 24 | 134 | 25 | 0 | 0 |
| May 24 | 136 | 23 | 0 | 0 |
| Jun 24 | 109 | 20 | 0 | 0 |
| Jul 24 | 151 | 24 | 0 | 0 |
| Aug 24 | 187 | 31 | 0 | 0 |
| Sep 24 | 195 | 27 | 0 | 0 |
| Oct 24 | 206 | 41 | 0 | 0 |
| Nov 24 | 178 | 35 | 0 | 0 |
| Dec 24 | 244 | 35 | 0 | 1 |
| Jan 25 | 400 | 196 | 0 | 0 |
| Feb 25 | 317 | 24 | 0 | 0 |
| Mar 25 | 214 | 19 | 1 | 0 |
| Apr 25 | 219 | 19 | 0 | 0 |
| May 25 | 164 | 28 | 0 | 0 |
| Jun 25 | 229 | 34 | 0 | 0 |
| Jul 25 | 177 | 29 | 0 | 0 |
| Aug 25 | 224 | 36 | 0 | 0 |
| Sep 25 | 205 | 19 | 0 | 0 |
| Oct 25 | 197 | 35 | 0 | 0 |
| Nov 25 | 197 | 19 | 1 | 0 |
| Dec 25 | 306 | 35 | 0 | 0 |
| Jan 26 | 236 | 20 | 1 | 0 |
| Feb 26 | 216 | 26 | 1 | 0 |
| Mar 26 | 197 | 20 | 1 | 0 |
| Apr 26 | 196 | 22 | 0 | 0 |
| May 26 | 223 | 27 | 0 | 0 |
| Jun 26 | 228 | 18 | 0 | 0 |
| Jul 26 | 203 | 25 | 0 | 0 |
| Aug 26 | 274 | 28 | 0 | 4 |
| Sep 26 | 197 | 32 | 0 | 2 |
| Oct 26 | 178 | 17 | 0 | 4 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Detachment of Device or Device Component585
- Device Slipped443
- Incorrect, Inadequate or Imprecise Result or Readings260
- Break214
- Device-Device Incompatibility178
- Malposition of Device148
- Material Deformation137
- Mechanical Problem134
- Use of Device Problem128
- Material Fragmentation122
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions2,618
- Bone Notching109
- Spinal Column Injury79
- Insufficient Information48
- Foreign Body In Patient30
- Pain21
- Laceration(s)11
- Nerve Damage11
- Unspecified Tissue Injury11
- Bone Fracture(s)10
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0119-2027 · Started Aug 28, 2026 · Posted by FDA Oct 7, 2026
Reason given: A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
Cause as FDA categorized it: Nonconforming Material/Component
- Think Surgical Square Driver REF: 109126Open, Classified
Z-0192-2027 · Started Aug 20, 2026 · Posted by FDA Oct 16, 2026
Reason given: Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.
Cause as FDA categorized it: Process control
- Velys Enabling Tech Robotic-Assist Station REF: 451611001Open, Classified
Z-3088-2026 · Started Jul 23, 2026 · Posted by FDA Sep 2, 2026
Reason given: Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
Cause as FDA categorized it: Under Investigation by firm
Z-2716-2026 · Started May 8, 2026 · Posted by FDA Jul 9, 2026
Reason given: Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
Cause as FDA categorized it: Process control
Z-2292-2026 · Started Apr 23, 2026 · Posted by FDA Jun 2, 2026
Reason given: The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
Cause as FDA categorized it: Device Design
Z-2293-2026 · Started Apr 23, 2026 · Posted by FDA Jun 2, 2026
Reason given: The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
Cause as FDA categorized it: Device Design
- Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee SystemOpen, Classified
Z-1629-2026 · Started Feb 10, 2026 · Posted by FDA Mar 24, 2026
Reason given: Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
Cause as FDA categorized it: Nonconforming Material/Component
- Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee SystemOpen, Classified
Z-1630-2026 · Started Feb 10, 2026 · Posted by FDA Mar 24, 2026
Reason given: Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
Cause as FDA categorized it: Nonconforming Material/Component
Showing the 8 newest of 115 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.