Device type · OOE
Ophthalmic Femtosecond Laser
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
OOE
Ophthalmic
Device class
Class II
FDA decisions since 2015
28
510(k), De Novo and PMA
Adverse event reports
420
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Ophthalmic Femtosecond Laser
- Product code
- OOE
- Device class
- Class II
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.4390
- Specialty
- Ophthalmology
- FDA record
- Classification database entry for OOE
Latest FDA decisions
The 8 newest of 28 since 2015.
K252682 · Decided Sep 24, 2025 · 30 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
K243896 · Decided Apr 28, 2025 · 131 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
Johnson & Johnson Surgical Vision, Inc.
K223566 · Decided Apr 14, 2023 · 136 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
K223838 · Decided Apr 12, 2023 · 111 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
K220259 · Decided Jun 9, 2022 · 129 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
K220516 · Decided May 17, 2022 · 83 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
Sie Ag,Surgical Instrument Engineering
K213559 · Decided Apr 21, 2022 · 168 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
K210701 · Decided Apr 7, 2021 · 29 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
Reports received
420
Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.
Malfunction reports
145
35% of reports
Injury reports
274
65% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 3 | 11 | 0 | 0 |
| Jan 24 | 33 | 9 | 0 | 0 |
| Feb 24 | 41 | 16 | 0 | 0 |
| Mar 24 | 23 | 13 | 0 | 0 |
| Apr 24 | 10 | 14 | 0 | 0 |
| May 24 | 10 | 16 | 0 | 0 |
| Jun 24 | 5 | 8 | 0 | 0 |
| Jul 24 | 14 | 16 | 0 | 0 |
| Aug 24 | 10 | 9 | 0 | 0 |
| Sep 24 | 9 | 13 | 0 | 0 |
| Oct 24 | 26 | 22 | 0 | 0 |
| Nov 24 | 9 | 14 | 0 | 0 |
| Dec 24 | 14 | 14 | 0 | 0 |
| Jan 25 | 16 | 8 | 0 | 0 |
| Feb 25 | 5 | 5 | 0 | 0 |
| Mar 25 | 1 | 15 | 0 | 0 |
| Apr 25 | 46 | 25 | 0 | 0 |
| May 25 | 5 | 19 | 0 | 0 |
| Jun 25 | 35 | 17 | 0 | 0 |
| Jul 25 | 19 | 19 | 0 | 0 |
| Aug 25 | 8 | 10 | 0 | 0 |
| Sep 25 | 9 | 70 | 0 | 0 |
| Oct 25 | 14 | 64 | 0 | 0 |
| Nov 25 | 13 | 27 | 0 | 0 |
| Dec 25 | 7 | 61 | 0 | 0 |
| Jan 26 | 11 | 60 | 0 | 0 |
| Feb 26 | 9 | 17 | 0 | 0 |
| Mar 26 | 8 | 15 | 0 | 0 |
| Apr 26 | 14 | 22 | 0 | 0 |
| May 26 | 5 | 13 | 0 | 0 |
| Jun 26 | 16 | 13 | 0 | 0 |
| Jul 26 | 20 | 20 | 0 | 0 |
| Aug 26 | 20 | 7 | 0 | 0 |
| Sep 26 | 12 | 11 | 0 | 1 |
| Oct 26 | 10 | 8 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem196
- Failure to Cut73
- Decrease in Suction38
- Unexpected Therapeutic Results22
- Patient Device Interaction Problem17
- Computer Software Problem14
- Appropriate Device Problem Term/Code Not Available13
- Failure to Align7
- Patient-Device Incompatibility7
- Mechanical Problem5
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Keratitis114
- Capsular Bag Tear82
- No Clinical Signs, Symptoms or Conditions80
- Insufficient Information79
- Corneal Perforation26
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available24
- Corneal Abrasion8
- Eye Injury8
- Visual Impairment7
- Unspecified Eye / Vision Problem2
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0444-2022 · Started Oct 21, 2021
Reason given: Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program...
Cause as FDA categorized it: Software design
Johnson & Johnson Surgical Vision Inc
Z-0066-2019 · Started May 9, 2018
Reason given: Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.
Cause as FDA categorized it: Device Design
SIE AG, Surgical Instrument Engineering
Z-1272-2018 · Started Jan 21, 2018
Reason given: This Correction is being initiated based on a customer complaint received on November 16, 2017 which indicated a complication with a cataract procedure. An investigation completed by the manufacturer determined that although the complication was due to user error improvements could and should be made on our part to...
Cause as FDA categorized it: Device Design
Z-1688-2016 · Started Apr 14, 2016
Reason given: Alcon is conducting a voluntary recall of all LenSx Lasers with Corneal Flap Capability after receiving reports concerning incomplete corneal flap creation, which manifests as an un-dissected area(s) within the side cut.
Cause as FDA categorized it: Software design (manufacturing process)
Z-1515-2015 · Started Feb 20, 2015 · Posted by FDA Apr 23, 2015
Reason given: Device may experience loss of suction during treatment resulting in scoring the cornea during lens fragmentation.
Cause as FDA categorized it: Device Design
Z-1683-2015 · Started Feb 20, 2015 · Posted by FDA May 28, 2015
Reason given: Software anomaly on OptiMedica CATALYS System version 3.00.05 that may result in incorrect parameters provided for cataract incision surgeon templates and both eyes have the same parameters.
Cause as FDA categorized it: Software design
Z-1445-2014 · Started Mar 14, 2014 · Posted by FDA Apr 9, 2014
Reason given: Alcon LenSx received reports of unexpected downward motion of the gantry.
Cause as FDA categorized it: Device Design
Z-0649-2014 · Started Dec 20, 2013 · Posted by FDA Jan 8, 2014
Reason given: Potential for unexpected heating of the transformer mounting bolt located within the system chassis and the subsequent melting of a non-slip computer shelf mat, which may manifest an odor and visible smoke.
Cause as FDA categorized it: Under Investigation by firm
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.