Device type · OTE
Digital Breast Tomosynthesis
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
OTE
Radiology
Device class
Class III
FDA decisions since 2015
2
510(k), De Novo and PMA
Adverse event reports
268
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Digital Breast Tomosynthesis
- Product code
- OTE
- Device class
- Class III
- Review panel
- Radiology
- Specialty
- Radiology & Imaging
- FDA record
- Classification database entry for OTE
Latest FDA decisions
Fujifilm Medical Systems U.S.A, Inc.
P160031 · Decided Jan 10, 2017 · 162 days after receipt
Device type: Digital breast tomosynthesis (OTE) · Radiology & Imaging
Siemens Medical Solutions USA, Inc.
P140011 · Decided Apr 21, 2015 · 299 days after receipt
Device type: Digital breast tomosynthesis (OTE) · Radiology & Imaging
Reports received
268
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
248
93% of reports
Injury reports
19
7.1% of reports
Death reports
1
0.4% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 8 | 0 | 0 | 0 |
| Jan 24 | 3 | 3 | 0 | 0 |
| Feb 24 | 11 | 0 | 0 | 0 |
| Mar 24 | 8 | 1 | 0 | 0 |
| Apr 24 | 5 | 0 | 1 | 0 |
| May 24 | 10 | 0 | 0 | 0 |
| Jun 24 | 14 | 1 | 0 | 0 |
| Jul 24 | 12 | 0 | 2 | 0 |
| Aug 24 | 13 | 0 | 1 | 0 |
| Sep 24 | 13 | 1 | 0 | 0 |
| Oct 24 | 6 | 0 | 0 | 0 |
| Nov 24 | 11 | 0 | 3 | 0 |
| Dec 24 | 11 | 0 | 0 | 0 |
| Jan 25 | 6 | 0 | 0 | 0 |
| Feb 25 | 7 | 0 | 1 | 0 |
| Mar 25 | 10 | 0 | 0 | 0 |
| Apr 25 | 20 | 0 | 0 | 0 |
| May 25 | 12 | 0 | 0 | 0 |
| Jun 25 | 10 | 0 | 0 | 0 |
| Jul 25 | 15 | 0 | 0 | 0 |
| Aug 25 | 9 | 0 | 0 | 0 |
| Sep 25 | 7 | 2 | 0 | 0 |
| Oct 25 | 14 | 1 | 0 | 0 |
| Nov 25 | 12 | 1 | 1 | 0 |
| Dec 25 | 13 | 1 | 0 | 0 |
| Jan 26 | 11 | 0 | 0 | 0 |
| Feb 26 | 13 | 0 | 0 | 0 |
| Mar 26 | 17 | 0 | 0 | 0 |
| Apr 26 | 15 | 2 | 0 | 0 |
| May 26 | 37 | 1 | 0 | 0 |
| Jun 26 | 32 | 1 | 0 | 0 |
| Jul 26 | 22 | 2 | 0 | 0 |
| Aug 26 | 23 | 3 | 0 | 0 |
| Sep 26 | 27 | 4 | 0 | 0 |
| Oct 26 | 26 | 4 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Hologic267
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Unintended System Motion140
- Unintended Movement39
- Mechanical Problem12
- Device Emits Odor10
- Vibration6
- Adverse Event Without Identified Device or Use Problem5
- Appropriate Device Problem Term/Code Not Available5
- Arcing5
- Mechanics Altered5
- Noise, Audible5
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions253
- Insufficient Information3
- Abdominal Pain1
- Ascites1
- Bone Fracture(s)1
- Breast Discomfort/Pain1
- Bruise/Contusion1
- Burn(s)1
- Cancer1
- Deformity/ Disfigurement1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Senographe Pristina 3DOpen, Classified
Z-0328-2027 · Started Aug 28, 2026 · Posted by FDA Nov 4, 2026
Reason given: GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the...
Cause as FDA categorized it: Radiation Control for Health and Safety Act
Z-2168-2026 · Started Apr 8, 2026 · Posted by FDA May 12, 2026
Reason given: Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
Cause as FDA categorized it: Device Design
FUJIFILM Healthcare Americas Corporation
Z-2219-2025 · Started Jun 11, 2025 · Posted by FDA Jul 30, 2025
Reason given: Devices had an unapproved slabbing software function enabled for use.
Cause as FDA categorized it: Process control
Z-3248-2024 · Started Aug 19, 2024 · Posted by FDA Sep 25, 2024
Reason given: Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis
Cause as FDA categorized it: Software design
Z-1144-2023 · Started Jan 16, 2023 · Posted by FDA Feb 21, 2023
Reason given: C-Arm unexpected movement may cause blunt trauma should the tube arm impinge upon an individual
Cause as FDA categorized it: Device Design
Siemens Medical Solutions USA, Inc
Z-0888-2016 · Started Feb 12, 2016
Reason given: A potential system fatal error may occur during tomosynthesis reconstruction with large breasts that cover nearly the entire detector surface and with a breast thickness larger than 90 mm. The data to be processed for the tomosynthesis application can exceed the capacity of the graphical processing unit (GPU). If the...
Cause as FDA categorized it: Software design
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.